3. If the control solution testing result is out of range:
POSSIBLE CAUSE
Error in performing the test.
Control solution vial was poorly
shaken.
Expired or contaminated control
solution.
Control solution that is too warm or
too cold.
Defective test strip.
Meter malfunction.
Malfunction of meter and test strip.

SYMBOL INFORMATION

SYMBOL
Consult instructions for
use
Temperature limitation
Use by
Batch code
Caution, consult
accompanying documents
REFERENT
WHAT TO DO
Read instructions thoroughly and
repeat the test again.
Shake the control solution vigorously
and repeat the test again.
Check the expiry date of the control
solution.
Control solution, meter, and test strips
should be at room temperature 20°C
to 25°C ( 68°F to 77°F) before testing.
Repeat the test with a new test strip.
Please contact GlucoRx Customer
care on 01483 755133.
Please contact GlucoRx Customer
care on 01483 755133.
SYMBOL
Manufacturer
Serial number
Collection for electrical
and electronic equipment
CE mark
Humidity limitation
REFERENT