Guidance and manufacturer's declaration - electromagnetic emissions
The Medel Professional is intended for use in the electromagnetic environment speci-
fied below. The customer or the user of the Medel Professional should assure that it is
used in such an environment
Emission test
Compliance
Emissions RF
Group 1
CISPR 11
RF Emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 62000-3-2
Voltage fluctuations/
Complies
flicker emissions
IEC 62000-3-
Guidance and manufacturer's declaration - electromagnetic emissions
The Medel Professional is intended for use in the electromagnetic environment specified below. The
customer or the user of the Medel Professional should assure that it is used in such an environment.
Immunity test
IEC 60601 Test level
Electrostatic
± 6kV contact
discharges
± 8kV air
IEC 62000-4-2
NOTE: UT is the a.c. mians voltage prior to application of the test level.
32
100024_MedelProfessional_USERmanual_REV.02.indd 32-33
Electromagnetic environment – guidance
The Medel Professional uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any interfe-
rence in nearby electronic equipment.
The Medel Professional is suitable for use in all
establishments other than domestic, and may
be used in domestic establishments and those
directly connected to the public low-voltage po-
wer supply network that supplies buildings used
for domestic purposes, provided the following
warning is heeded:
Warning: This equipment/system is intended
for use by healthcare professionals only. This
equipment/ system may cause radio interfe-
rence or may disrupt the operation of nearby
equipment. It may be necessary to take mitigation
measures, such as re-orienting or relocating the
Medel Professional or shielding the location.
Compliance level
Electromagnetic
environment - guidance
± 6kV contact
Floors should be wood,
concrete or ceramic tile.
± 8kV air
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Guidance and manufacturer's declaration - electromagnetic emissions
The Medel Professional is intended for use in the electromagnetic environment specified below. The
customer or the user of the Medel Professional should assure that it is used in such an environment.
Electrical fast
±2 kV for power supply
transient/burst
lines
IEC 62000-4-4
±1 kV for input/output lines
Surge
±1 kV line(s) to line(s)
IEC 62000-4-5
±2 kV line(s) to earth
Voltage dips,
<5% UT
short interrup-
(>95% dip in UT)
tions and voltage
for 0.5 cycle
variations on
40% UT
power supply
(60% dip in UT)
input lines
for 5 cycle
70% UT
IEC 62000-4-11
(30% dip in UT)
for 25 cycle
<5% UT
(>95% dip in UT)
for 5s
Power frequency
3 A/m
0/60Hz)
magnetic field
IEC 62000-4-8
NOTE: UT is the a.c. mians voltage prior to application of the test level.
±2 kV for power
Mains power quality
supply lines
should be that of a typical
commercial or hospital
environment.
The electrical fast transient
burst (EFT) is generated by
the switching of inductive
loads. Separation between
the equipment and other
loads shall be considered
before installation.
Mains filter is required, if
necessary.
±1 kV line(s) to
Mains power quality
line(s)
should be that of a typical
commercial or hospital
environment.
5% UT
Mains power quality
(>95% dip in UT)
should be that of a typical
for 0.5 cycle
commercial or hospital
40% UT
environment. If the user
(60% dip in UT)
of the Medel Professional
for 5 cycle
requires continued
70% UT
operation during power
(30% dip in UT)
mains interruptions, it is
for 25 cycle
recommended the Medel
<5% UT
Professional be powered
(>95% dip in UT)
from an uninterruptible
for 5s
power supply or a battery.
3 A/m
Power frequency magnetic
fields should be at
levels characteristic of a
typical location in a typical
commercial or hospital
environment.