Supplement Caster Brakes to Instructions for Use
YOU NEED ONE COPY PER INSTRUCTIONS FOR USE. PLEASE COPY IF NECESSARY.
EN / AU
Supplement to Dialog
brakes
Read the following information and instruct all users in your organization. Add this supplement
to your Instructions for Use.
This supplement only contains information that are changed. For the complete descriptions,
refer to the existing IFU Dialog
Chapter 3.3.1 Wheeling (page 3-4)
Risk of damage if dialysis machine is tilted by > 10°!
Have two or more persons at hand for transporting the machine on stairs and inclined areas.
Do not tilt the dialysis machine by more than 10°.
CAUTION
Fig. 3-1
Release the brakes from all casters.
Wheel the dialysis machine.
Reapply the brakes to all casters.
Supplement 38910480 / Rev. 2.88.00 / December 2018
+
Instructions for Use (IFU): Information about caster with caster
+
9.1x, rev. 2.17.xx.
Transport on stairs and slopes (2 persons)
+
Dialog
1