Set the SpO
alarm limits as follows:
2
Choose SpO
Function button to enter setup. The SpO
box will change from real line box to broken line box.
The adjustable range for upper limit of SpO
"1 + the lower limit of SpO
SpO
default value after the oximeter is turned on again.
Press the Up Arrow or Down Arrow button to
increase or decrease values.
Choose SpO
Function button to set up it. The SpO
will change from real line box to broken line box. The
adjustable range for the lower limit of SpO
from 0 to "the upper limit of SpO
value of SpO2 Lo Alarm is set to less than 85, it will
restore to 85 after the oximeter is turned on again.
Press the Up Arrow or Down Arrow button to
increase or decrease values.
SpO
by at least 1%.
Press the Function button, confirm the alarm range
setup.
Press the On/Off button to exit the menu, and return to
measurement state.
3.3.10 PR Alarm Setup
The user can use PR Hi Alarm and PR Lo Alarm in menu to
adjust pulse rate alarm limits.
By default, PR Hi Alarm and PR Lo Alarm are 170 and 40
respectively.
Set the PR limits as follows:
Choose PR Hi Alarm in the menu, press the Function
button to enter setup. The PR Hi Alarm box changes
from real line to broken line. The adjustable range of
Hi Alarm in the menu, press the
2
Hi Alarm is set to less than 85, it will restore to
2
Lo Alarm in the menu, press the
2
Hi Alarm is always higher than SpO
2
" to 100. If the value of
2
35
Hi Alarm
2
is from
2
Lo Alarm box
2
Alarm is
2
Alarm - 1". If the
2
Lo Alarm
2