Electronic Digital Thermometer adopts scientific technology of microcomputer. It breaks through the traditional disadvantages of the
mercury thermometer of mercury pollution and inconvenient reading with features of fast measurement, safety and simply operation.
Specification
Measuring Range:
32.0°C~43.9°C (90°F~111°F)
Measuring time:
About 10 seconds
Accuracy:
±0.1°C (±0.2°F) for 35.0~41.0°C (95°F~105.8°F) & ±0.2°C
(±0.4°F) for remaining range at an ambient of 22°C (71.6°F)
Display:
31/2 LCD digits (Unit of display: 0.1°C/0.1°F)
Battery:
CR2032 lithium
Battery life:
About 1000 times
Weight:
27g
Thermometric sensor
How to use
1) Please make sure the environment temperature is between 16°C to 31°C before use.
2) Press power switch, the screen shows the full segment with a sound of "Bi", which indicates sufficient
power.
3) Memory function: After the full segment, the last measured value with the letter of "Last memory" are
displayed.
4) After 2S, it will displays "Lo°C" or "Lo°F". Then the symbol "°C" or "°F" will flash. Please wait for the
"°C" or "°F" to flash 3 times before taking measurement.
5) A buzzing tone will be heard for about 3 seconds which means the finish of measurement.
a. Normal alarm: sound of "bibi.........bibi.........bibi........." with green backlight 5 seconds means normal temperature that
below 37.6°C/99.7°F.
b. Fever alarm: Sound of "bibibibi...bibibibi...bibibibi..."with red backlight 5 seconds means fever that the temperature is above
37.6°C/99.7°F.
6) During measurement: LCD will display "Lo°C" or "Lo°F" if the measured temperature is below 32.0°C (90°F), or will display "Hi°C"
or "Hi°F" if the measures temperature is above 43.9°C/111°F.
7) Turn off the unit by just slightly pressing power switch, or leave it for a few minutes, it will auto-off to save power.
8) When it is used to measure children temperature, close supervision shall be provided by an adult.
9) To maintain the accuracy of each measurement, please wait for at least 3 minutes before another measurement is taken.
10) How to change "°C" to "°F": Press switch to turn on and hold it, until displays "Lo°C" (no flash), then release the switch within 2S
and press it again, you can change the temperature unit from °C to °F or back.
Measurement:
You are advised to measure your temperature at specific time in morning, if it is earlier or later, the temperature would be lower or
higher.
1) Used in mouth: Please close your mouth for 1 minute for stability of temperature, breath calmly through nose.
Respiratory air would affect the accuracy of measure of your mouth.
2) Used in anus of rectum: This is the way that can get the most steady temperature and suits for both infants and children. The
sensor shall be inserted for 2~3 cm. Vaseline can be used for easier insertion.
3) Used under armpit: Clean the thermometer with a dry cloth and put it under armpit. Press and fix it with the arm. The temperature
value of this measurement is not so accurate that not recommended for accurate measurement.
The normal temperatures are different from the different human bodies and measuring time.
Cleaning is required before every measurement.
Measurement is inaccurate when the battery is low-power
Replace Battery
1) When the display becomes indistinct or when "
measurement.
2) To replace the battery, open the cover of the end part according to the indication with a coin and takeout the used battery.
3) When replace with a new battery, make sure to use the new CR2032, then re-insert the cover with the coin, avoid damage to the
waterproof ring.
LCD display
Switch
" appears on LCD, the battery must be replaced at once for accurate
ENGLISH
Cleaning and maintenance:
The thermometer is waterproof and can be cleaned by water or cleaning lotion with soft cloth: Cleaning lotion below is recommended:
Name
Density
Medical alcohol
75%
Notes: Other lotion, density or cleaning method may damage the thermometer. The product doesn't need any calibration
and is recommended to use for 3 years.
Remarks:
1) It is suitable for measuring temperature of human body and necessary of families.
2) It is designed and manufactured complying with relevant requirements in EN12470-3: Performance of Compact electrical ther-
mometers (non-predictive and predictive) with maximum device.
3) It has passed EMC emission test, immunity test according to EN60601-1-2 and biocompatibility test according to MDD annex I.
4)
According to the degree of protection against electric shock, it is classified to BF, in accordance with EN60601-1.
5) Avoid falling, crash or knock.
6) Never put it in boiled water for sterilizing.
7) Never touch any chemical solution and keep away from straight sunlight or heat source as well as strong electromagnetic source
to avoid interference.
8) It shall be operated in an environment at the temperature of 10°C~35°C (50°F~95°F) and the humidity of 40~80% while stored
and transported in an environment at the temperature of 0°C~40°C (32°F~104°F) and humidity of 30~90%.
9) Don't remove any parts except replacing battery.
10) Please replace battery before claim for replacing a new product.
Full segment
11) The scrap of product and battery are to be disposed according to relevant national regulations.
12) Caution! Be sure that both the battery and the thermometer are kept away from children.
13) To prevent battery leakage corrosion, please remove the battery if no use long time.
14) The product doesn't need any calibration and is recommended to use for 3 years.
15) Children should be under close supervision at all times while using this product. After monitoring the temperature, the unit should
be removed from the child immediately.
Product code
Humidity limit
%
Keep in a cool, dry place
Medical Device complies
with Directive 93/42/EEC
Authorized representative
in the European community
Disposal: The product must not be disposed of along with other domestic waste. The users must dispose of this equipment
by bringing it to a specific recycling point for electric and electronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
25553 / KD-602
YanCheng KellyUnion Import and Export Co., LTD
Room 2509, Building 1, Financial City, No.5,
Shiji Avenue, Chengnan New District,
224005 Yancheng City, P.R. China - Made in China
Importato da / Imported by:
Gima S.p.A.
Via Marconi, 1 - 20060 Gessate (MI) Italy
gima@gimaitaly.com - export@gimaitaly.com
www.gimaitaly.com
Battery cover
CR2032 Battery
Lot number
Manufacturer
Temperature limit
Date of manufacture
Keep away from sunlight
Type BF applied part
Caution: read instructions
Follow instructions for use
(warnings) carefully
WEEE disposal
PROLINX GMBH
Brehmstr. 56, 40239, Duesseldorf
40°C
90%
0°C
30%