Descargar Imprimir esta página

Aglient Turbo-V 70 PCB Manual De Instrucciones página 268

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 55
17
Technical Information
Error Messages
After the starting phase if the output connection is shorted (output current
NOTE
higher than 2.2 A), the display will be as shown in the following figure.
F
S
Check connections and shortages between pump and controller, then
press the STOP RESET pushbutton twice to start the pump.
If the pump is stopped by an emergency stop signal provided via a remote
NOTE
contact, the display will be as shown in the following figure.
Remove the controller power cable and check the emergency
condition. Then reconnect the power cable and press the START
pushbutton to start the pump.
The following message is displayed if the fuse of the transformer
secondary blows:
Since no operation can be carried out, contact AGILENT.
268/280
A
U
L
T
H
O
R
T
S
O
V
C
O
N
F
A
Turbo-V 70 PCB Controller User Manual / 87-900-871-01 (E)
:
C
I
R
C
Y
S
T
E
M
E
R
R
I
D
T
R
O
I
L
U
R
U
I
T
E
L
L
E
R
E

Publicidad

loading

Este manual también es adecuado para:

969-9508