Display blank after turn on
test, Error message, Display is
locked
OL or ---- appears on display
---- or NULL displayed
Display is unstable
Weight value incorrect
Cannot use Full Capacity
Not linear
Off center
loading error
Battery will not charge
11. WARRANTY
This scale is warranted against defects of manufacturing and materials for a period of 1 year, from the delivery date.
During this period, GIROPÈS SL will take charge of repairing the scale
This warranty does not cover defects or damaged caused by misuse, overloading or improper installation contrary to the recom-
mendations described in this manual.
This warranty does not cover shipping costs for the reparation of the balance.
Pan not installed.
Unstable weight.
Load cell damaged.
Mechanics damaged.
Maximum capacity exceeded. Load cell
or mechanics damaged. Power supply
faulty.
Weight on scale below permissible
li-mit.
Pan has been removed.
Pan support not seated properly. Power
supply faulty.
Load cell or mechanics damaged.
Drafts or air currents.
Obstruction under pan.
Sample is moving (animal weighing).
Vibrations through table. Temperature
changed dramatically. Power supply
faulty.
Calibration error, Recalibrate. Linearity
error, set Linearity.
Unit calibrated with inaccurate weight.
Balance not level.
Obstruction between sample and
co-ver.
Wrong unit of weight displayed.
Over load Stops hitting pan support or
hitting bottom of load cell.
Shipping screw not removed if
appli-cable.
Electronic problem on A/D. Parameters
set incorrectly.
Load cell or mechanics damaged.
Overload stops hitting too soon.
Load cell or mechanis damaged.
A/D damaged.
Adjust mechanics.
Overload stops not correct.
Load cell damaged.
Mains voltage not present or too low.
Charging circuit failure.
Battery failure.
USER MANUAL BAR
Check the pans are installed correctly.
Try turning the scale on again.
Check the pans are installed correctly.
Try turning the scale on again.
Check the pans are installed correctly
Try pressing
.
Try turning the scale on again.
Verify the scale is in a acceptable loca-
tion and on a good table.
Verify the power supply is correct for
the scale.
Calibrate again, paying special atten-
tion to the mass used, the stability of the
scale, and the weighing units re-quired.
If linearity can be set by the user it
will be described in the user manual.
Check pan is installed correctly. Verify
installation is acceptable.
Look for obstruction under pan, ship-
ping screws, and verify pan installa-tion.
Check the weighing units used.
Look for obstruction under pan, ship-
ping screws, and verify pan installa-tion.
If linearity can be set by the user it will
be described in the user manual.
Look for obstruction under pan, ship-
ping screws, and verify pan installa-tion.
Verify the batteries are rechargeable
types.
Check power supply voltage is correct.