Maintenance And Cleaning; Warranty; Technical Data - IKA Vortex 2 Manual De Instrucciones

Tabla de contenido

Publicidad

The instrument is maintenance-free. It is only subject to the natu-
ral wear and tear of components and their statistical failure rate.
Cleaning
Use only cleaning agents which have been approved by IKA to
clean IKA instruments.
Dirt
Cleaning agent
Dyes
isopropyl alcohol
Construction materials water containing tenside/isopropyl alcohol
Cosmetics
water containing tenside/isopropyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside
For materials which are not listed, please request information from
IKA application support.
Wear protective gloves during cleaning the instruments.
Electrical instruments may not be placed in the cleansing agent for
the purpose of cleaning.
Do not allow moisture to get into the instrument when cleaning.
Before using another than the recommended method for cleaning
In accordance with IKA warranty conditions, the warranty period
is 24 months. For claims under the warranty please contact your
local dealer. You may also send the machine direct to our factory,
enclosing the delivery invoice and giving reasons for the claim. You
will be liable for freight costs.
Operating voltage / Design frequency
or
or
Input power
Speed
Speed adjustment
Speed display
Drive
Agitation stroke
Shaking motion
Perm. duration of operation
Perm. ambient temperature
Perm. relative humidity
Protection type (acc. to DIN EN 60529)
Protection class
Overvoltage category
Contamination level
Operation at a terrestrial altitude
Dimensions total (W x D x H)
(without attachment)
Weight without supported load
Max. supported load
(incl. attachment)
Subject to technical changes!

Maintenance and Cleaning

For cleaning disconnect the main
plug!

Technical data

V/Hz
V/Hz
V/Hz
W
rpm
mm
%
°C
%
m
mm
kg
kg
or decontamination, the user must ascertain with IKA that this
method does not destroy the instrument.
The rubber of the attachments' pads (VG 3.2 and VG 3.3) is a
natural product, so talcum can discharge. Clean the rubber pad
with glycerin.
Spare parts order
When ordering spare parts, please give:
- Machine type;
- Manufacturing number, see nameplate;
- Item number and designation of the spare part,
see www.ika.com.
Repair
Please send instrument in for repair only after it has been
cleaned and is free from any materials which may consti-
tute a health hazard.
For this you should request the "Decontamination Clearance
Certificate" from IKA, or use the download printout of it from the
IKA website www.ika.com.
If you require servicing, return the instrument in its original pack-
aging. Storage packaging is not sufficient. Please also use suitable
transport packaging.

Warranty

The warranty does not cover wearing parts, nor does it apply to
faults resulting from improper use or insufficient care and mainte-
nance contrary to the instructions in this operating manual.
220...240 ±10% / 50/60
115±10%
/ 60
100±10%
/ 50/60
60
(750 - 2750) ± 250, at 50 Hz, infinitely adjustable
(750 - 3300) ± 250, at 60 Hz, infinitely adjustable
analog (infinitely)
rotating knob in front of the device
Scale 0 - 6
Shaded pole motor
4
horizontal, approximately circular
100
+5 to +40
80
IP 21
(IP 21 can only be achieved when fitted with top attachment)
I (protective earth)
II
2
max. 2000
120 x 138 x 140
3.9
0.4
12

Publicidad

Tabla de contenido
loading

Tabla de contenido