Err 8.4
The pan is not properly installed.
Err 8.5
The tare value is a negative value.
Err 9.5
System error
Err 13
Lo.Bat
Battery very low
The value is out of zero range.
--NO--
The zero range is dependent on
region.
------
The environment is not stable.

6.3 Service Information

If the troubleshooting section does not resolve or describe your problem, contact
your authorized OHAUS service agent. For service assistance or technical support
in the United States call toll-free 1-800-672-7722 ext. 7852 between 8:00 AM and
5:00 PM EST. An OHAUS product service specialist will be available to provide
assistance Monday through Friday. Outside the United States, Canada and Puerto
Rico, please visit our web site, www.ohaus.com to locate the OHAUS office nearest
you.

7. TECHNICAL DATA

Equipment Ratings:
Indoor use only
Altitude: 2000m
Operating temperature: 0 to 40° C
Humidity: Maximum relative humidity 80% for temperatures up to 31 ° C decreasing
linearly to 50% relative humidity at 40° C.
Electrical supply: 12VDC, 0.5A. For use with certified or approved power supply,
which must have a SELV and limited energy circuit output.
Voltage fluctuations: Mains supply voltage fluctuations up to ± 10% of the nominal
voltage.
Overvoltage category (Installation category): II
Pollution degree: 2
Install the weiging pan properly.
Press Zero instead of Tare.
Disconnect the power and
restart.
If error persists, please contact
OHAUS service
(www.ohaus.com - More -
Contact Us)
Replace dry batteries or charge
rechargeable battery.
Perform zero again according to
regional requirement.
Move balance to a suitable
location.
EN-15