Pump operations

Mains Connection

It is recommended to connect the pump to the power supply through a dedicated
CAUTION
Thermal magnetic circuit breaker (protection against short-circuit and line
overload).

Oil and Filter Cartridge Replacement Procedures

Oil Change Procedure

Remove the inlet and outlet pipes before opening the oil drain valve.
WARNING
1
Oil drain valve
Figure 25 Oil Change
MS40-S User Manual
Place a tank for waste oil under the oil drain plug.
Fill oil cap