Error codes

If an error condition occurs, win206 assumes a value other than 0 and precisely:
If more than one error occurs at the same time, the error code displayed will be the
sum of the codes of the individual errors.
MS40-S User Manual
N
R
N
R
N
R
N
R
A
R
A
R
N
RW
L
RW
A
R
N
RW
L
RW
128: pump rotor blocked
64: short circuit on the motor circuit
32: the mains voltage was out of specification during pump operation
16: motor overtempt
8: the mains voltage was too low (out of specification) when the pump was
started
4: controller over temperature
2: internal error (internal DC-Bus Undervoltage)
1: motor overcurrent
LAST WORKING HOUR AVERAGE ELECTRICAL POWER TO MOTOR [W]
LAST WORKING HOUR AVERAGE FREQUENCY [Hz]
AVERAGE TEMPERATURE [°C]
AVERAGE POWER DELIVERED TO THE MOTOR [W]
FIRMWARE PROGRAM LISTING CODE & REVISION
FIRMWARE PARAMETER LISTING CODE & REVISION
RS485 SERIAL ADDRESS SETTING [0÷31] 0
SERIAL COMMUNICATION SELECT (0=RS232; 1=RS485)
PARAMETER LISTING CODE & REVISION
RS485 SERIAL ADDRESS SETTING [0-31]
SERIAL TYPE SELECT (0=RS232; 1=RS485)