Electromagnetic Immunity
Immunity Test
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
Power
Frequency
(50/60/400 Hz)
Magnetic Field
IEC 61000-4-8
Radiated RF
IEC 61000-4-3
All manuals and user guides at all-guides.com
Regulatory Information
IEC 60601
Compliance
Test Level
Level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
3 A/m
3 A/m
3 V/m
3 V/m
80 MHz to
2.5 GHz
35
Electromagnetic
Environment -
Guidance
There are no special
requirements with
respect to electrostatic
discharge.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial/
hospital environment.
There are no special
requirements for
non-commercial/
non-hospital
environments.
Portable and mobile
RF communications
equipment should
be used no closer
to any part of the
CPRmeter 2, than is
absolutely necessary.
†,‡ The recommended
separation distances
for various transmitters
and the CPRmeter
2 are shown in
the following table.
Interference may
occur in the vicinity
of equipment marked
with the following
symbol: