Airflo Plus
Battery Charging - fig.
The Airflo Plus is fitted with a quick release battery
pack. The batteries can either be re-charged whilst
in the Airflo Plus or easily removed and re-charged
in the optional charging cradle.
• Check the battery charger casing and lead for
signs of damage and if damaged DO NOT USE.
• Fit the appropriate mains plug adaptor to the
battery charger (A).
• Plug the battery charger into a mains outlet
socket and connect the charger output lead (B)
to the Airflo Plus charging socket (C).
• The battery status indicator light (D) will flash
green while the battery is charging and then
turn to a steady (non flashing) light when the
Airflo Plus is ready for use.
NOTE: The charger is not capable of providing
sufficient power to charge the batteries while the
compressor is running.
To maintain battery condition, It is recommended
that the Airflo Plus is left continuously on
charge when not in use. The batteries cannot be
overcharged using this charger.
WARNING: DO NOT recharge the batteries in a
bathroom or in a damp environment.
To disconnect the charger from the mains power
supply, unplug the charger from the mains power
socket.
4
3
B
3
A
C
D