Compressor does not run when the 4 way
hand control buttons are pressed.
Ensure that the Airflo Plus has been switched
on.
Check that there are no kinks in the air hose.
Compressor keeps stopping when trying to
inflate the lifting aid.
Check that there are no kinks in the air hose.
Airflo Plus
Weight:
Output pressure:
Batteries:
Battery Life:
Lifting performance (from full battery):
Environmental protection: IP51 (Protection against large dust deposits & vertical falling drops of water)
Type BF equipment
This equipment has been tested and meets the requirements of current EMC legislation
Charger: Model number - CA0431
Input:
Output:
Double insulated Class II equipment
Simple solutions for everyday independence
Trouble shooting guide
Specifications
5.7Kg (12½ lbs)
0.35 bar (5 psi)
2 x 12v - 2.3Ah (Maintenance free Valve Regulated Sealed Lead Acid)
Approx 1000 charging cycles (2 - 5yrs Depending on usage)
ELK - 3 lifts
Camel - 1 lift (minimum)
AC 100v - 240v, 50-60Hz, 0.3A
DC 30v, 0.45A
Battery State Indicator light flashing 'Red'.
This happens when the compressor's internal
temperature cut-out has tripped. Leave the
Airflo Plus for 60-90 minutes to cool.
Battery State Indicator light flashing
Green'.
This happens when the current overload cut-
out has tripped. Leave for 3 seconds to self
reset.
Compressor causes electrical interference
with another piece of electrical equipment.
Move either the compressor or other item
further away from each other until the
interference stops.
Airflo Plus
'Red/
7