Cleaning / Disinfection
Clean on a regular basis. Clean surfaces with a damp cloth using
an appropriate pH-neutral detergent. Do not use solvents or
strong liquids, this may damage surfaces on the lifter. For disinfec-
tion when needed; use isopropyl alcohol. Avoid abrasive cleaning
products. Check emergency stop and emergency lowering after
cleaning. The lift should not be exposed to running water.
Make sure not to damage or remove labels
when cleaning.
Reconditioning
Follow cleaning and installation procedure, complete periodic
inspection and use checklist after installation to recondition the
lifter.
Recycling
Refer to "Recycling instructions" for how to properly dispose of
product. This can be found on www.etac.com.
Spare parts
A list of spare parts is available on request.
Periodic inspection scope
Periodic inspection is a visual examination (particularly of the
lifter's load bearing structure and lifting mechanism with
attachments, brakes, controls, safety devices and person-support
devices) according to Periodic Inspection Report for Molift Air
200 and 350.
See www.etac.com.
If there is any problems with the lifter that
could jeopardize someones safety, the lifter
shall immediately be taken out of service and
marked "out of order". Do not use the lifter
untill it is repaired.
Periodic Inspection shall be performed at least once a year or
more frequently if required by local requirements. The inspection
must be performed by service personnel authorized by Etac.
Contact Etac at molift@etac.com for training and
authorization or recommendation of an approved service partner.
When performing a periodic inspection, the inspector shall fill
out the inspection report for Molift Air. The reports should be
retained by the person(s) responsible for servicing the lifter. If the
inspection reveals defects and damages, the lifter shall be taken
out of operation and marked "out of order" until it is repaired.
All manuals and user guides at all-guides.com
Troubleshooting
8
Symptom
Possible Cause/Action
The lifter does not respond
Emergency Stop is activated.
to Hand control action
Deactivate by pushing button
back in.
The lifters electronics is
overheated.
Wait for it to cool down.
Hand control is not plugged
in properly. Hand control or
plug or cord can be broken an
should be replaced.
The lifter does not respond
Lifting strap angle sensor is
to Hand control up button
activated. Adjust lifting strap
or move hoist to reduce lifting
strap angle.
When using Electrical
Lifting band has gone
Emergency lowering the
completely out, and has been
lifting band goes upwards
winded the wrong way.
Use manual emergency lower-
ing. Send lifter to service.
Molift Air / www.etac.com