10 A - English
Emissions and immunity information
The handle and charging accessory is intended for use in the electromagnetic environment specified below.
The customer or user of the handle and charging accessory should assure that it is used in such an
environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
The handle and charging accessory is intended for use in the electromagnetic environment specified below.
The customer or the user of the handle and charging accessory should assure that it is used in such an
environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Welch Allyn 719 Series Lithium Ion Handle and USB Charging Accessory
Electromagnetic emissions
Compliance
Electromagnetic environment - guidance
Group 1
The handle and charging accessory uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
Class B
The handle and charging accessory is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low voltage power supply
network that supplies buildings used for domestic purposes.
Class A
Complies
Electromagnetic immunity
IEC 60601 test level Compliance level
±8 kV contact
±8 kV
±15 kV air
±15 kV
±2 kV for power
±2 kV
supply lines
±1 kV for input/
±1 kV
output lines
±0.5 kV, ±1 kV
±1 kV
Line- to -line
±0.5 kV, ±1 kV, ±2 kV ±2 kV
WARNING This equipment/system is
intended for use by healthcare professionals
only. This equipment/ system may cause radio
interference or may disrupt the operation of
nearby equipment. It may be necessary to take
mitigation measures, such as re-orienting or
relocating the handle and charging accessory
or shielding the location.
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.