Warning: prongs and masks:
completely into the pro gsào à asksàa dàdo 'tàuseà
Always insert the Miniflow
®
creme near the prong, mask or Miniflow
®
slip of the Miniflow
.
Do not re-use the prongs or masks. They are for single patient use only.
Sterilized and disinfected prongs and masks may harden and cause injury to
the patient.
(5)
Select a bonnet of the correct size using the measuring tape and the
corresponding colour coding. The bonnet should be secured firmly
enough but without leading to deformation of the head. Remove the
separate Velcro strap from the bonnet and put it aside, within easy
reach. The bonnet is placed on the head such that the gap between
the two fleece parts is centred over the forehead. The bonnet should
cover the ears and be pulled down to the nape of the neck. Thread the
fixation straps through the loops on the prong or on the mask (see
picture).
In contrast to the disposable bonnet, the reusable bonnet is equipped
with a Velcro closure in which the interface is secured. The tubes are
di e tedà ehi dàtheàpatie t'sàhead.
Instruction: bonnets
Pullàtheà o
etàdo
àtoàjustàa o eàtheàpatie t'sàe e o s,àasà loseàasà
possible to the eyes – this makes it easier to secure the Medijet
Bonnets which are too s allàleadàtoàdefo
Bonnets which are too large cannot ensure sufficient fixation of the CPAP
generator.
(6)
Select the most appropriate angle for the prong adapter of the
Miniflow
®
(45° or 60°). If a prong is used, it generally yields the
narrower angle; if a mask is used, it generally yields the wider angle.
Place the Minifoam foam wedge between the two fleece pieces in the
middle of the bonnet. Fit the Miniflow
foam provided. Insert the prong into the nostrils. In doing so, maintain
a small distance between the nasal septum and the base of the prong.
When using the mask, ensure that it rests evenly on the skin. The
nostrils must be kept clear and should not be displaced by any part of
the mask. Using the Velcro strap, secure the Miniflow
the fleece pieces of the bonnet. Secure the Velcro area of the strap in
the area of the ears as an extension of the fixation loop of the prong.
Do not pull the straps too tightly; they should only provide the lateral
support needed.
Important instruction: prong and mask application
Do not insert the prongs fully into the nostrils. The connector
block should not touch the nose.
Regularly switching between prong and mask is an effective
ethodàfo àa oidi gào à edu i gàda ageàtoàtheàpatie t'sàski .à
Do not pull the straps too tight. The Miniflow
by means of the bonnet and the foam. If the leak is too large,
check the selected prong or mask size – pulling the straps more
tightly does not by itself solve the problem.
Excessively tightening the straps or inserting the prongs too far
can lead to pressure points on the baby's nose or skin.