5. Commissioning, operation
Permissible ambient and medium temperatures remain within the performance limits. Consider possible restrictions
■
on the ambient temperature range caused by mating connector used.
→ For performance limits see chapter 9 "Specifications"
5.3 Mechanical mounting
EN
The max. torque depends on the mounting point (e.g. material and shape). If you have any questions,
please contact our application consultant.
→ For contact details see chapter 1 "General information" or the back page of the operating instructions.
1. Seal the sealing face (→ see "Sealing variants").
2
At the mounting point, screw the pressure sensor in hand-tight.
.
3
Tighten with a torque spanner using the spanner flats.
.
Sealing variants
Parallel threads
Seal the sealing face with flat gasket, lens-type
sealing ring or WIKA profile sealing.
Tapered threads
Wrap threads with sealing material
(e.g. PTFE tape).
10
All manuals and user guides at all-guides.com
per EN 837
NPT, R and PT
WIKA operating instructions, pressure sensor, model A-1200
per ISO 1179-2 (formerly DIN 3852-E)