SAVE THESE INSTRUCTIONS

INTRODUCTION

This manual provides information necessary to operate the EVOLUTION
conserver can be used with CGA 870 post-valve cylinder [see Fig. A] at home or away from home to provide your
specific oxygen requirements. It requires two (2) 1.5 volt AA alkaline batteries for operation.
PURPOSE:
TM
The EVOLUTION OM-900CE electronic oxygen conserver includes a combination of a low-pressure regulator and an oxygen
conserver. It is designed for use with a cylinder as an ambulatory oxygen system and is capable of delivering a specific
amount of supplemental oxygen at the optimal point in the breathing cycle. The EVOLUTION increases the efficiency in the
delivery of oxygen, maximizing the beneficial effects and eliminating unnecessary oxygen waste.
FUNCTION :
When we breathe, approximately one-third of the time is spent inhaling and
two-thirds exhaling. As a result, oxygen delivered by continuous flow is
wasted during exhalation. By eliminating oxygen flow during exhalation, a
two-thirds saving is possible. Additionally, the oxygen available during the
very first part of inhalation contributes most to meeting oxygen needs.
The OM-900 CE electronic oxygen conserver takes advantage of these facts
to provide maximum efficiency in the delivery of oxygen. This device is
designed to be an integral component of a lightweight, long-lasting
supplemental ambulatory oxygen system.
TM
OM- 9 00 CE electronic oxygen conserver.This
TM
Toggle
Parallel Holes
on the
Post Valve
F IGURE A
Post-Valve Cylinder
7