Guidance of Manufacturer's Declaration-Electromagnetic Emissions
The model JC-114N (MH-DR103) is intended for use in the electromagnetic environment specified below. The customer or the user of the model JC-
114N (MH-DR103) should assure that it is used in such an environment
eMIssIoN TesT
Rf emissions
CIsPR 11
Rf emissions
CIsPR 11
Harmonic emissions
IeC61000-3-2
voltage fluctuation/
flicker emissions
IeC61000-3-2
Guidance of Manufacturer's Declaration-Electromagnetic Immunity
The model JC-114N (MH-DR103) is intended for use in the electromagnetic environment specified below. The customer or the user of the model JC-
114N (MH-DR103) should assure that it is used in such an environment
IMMuNITy TesT
IeC60601 TesT level
electrostatic discharge
±6kv contact
IeC61000-4-2
±8kv air
±2kv for power supply
electrical fast transient /
lines
burst IeC61000-4-4
±1kv for
Input /output lines
±1kv line(s) to l
surge IeC 61000-4-5
±2kv line(s)to e
< 5% ur (>95% dip in ur)
voltage dips,
for 0.5cycle
interruption and voltage
40% ur (>65% dip in ur)
short variations on
for 5 cycle
power supply input line
70% ur (>30% dip in ur)
IeC 61000-4-11
for 25cycle
< 5% ur (>95% dip in
ur) for 5 s
All manuals and user guides at all-guides.com
CoMPlIaNCe
eleCTRoMaGNeTIC eNvIRoNMeNT- GuIDaNCe
The model JC-114N (MH-DR103) uses Rf energy only for its internal
Group 1
function. therefore, its Rf emissions are very low and are not likely to cause
any interference in nearby electronic equipment
Class b
The model JC-114N (MH-DR103) is suitable for use in all establishments
including domestic establishments its and those directly connected to the
Class a
public low power supply network that supplies buildings used for domestic
purposes
Compliance
CoMPlIaNCe level
±6kv contact
±8kv air
±2kv for power supply
lines
±1kv differential mode
< 5% ur
(>95% dip in ur) for
0.5cycle
40% ur (>65% dip in
ur) for 5 cycle
70% ur (>30% dip in ur)
for 25 cycle
< 5% ur (>95% dip in
ur) for 5 s
45
eleCTRoMaGNeTIC eNvIRoNMeNT-GuIDaNCe
floor should be wood, concerts or ceramic tile. If floors are
covered with synthetic material ,the RH should be 30%
Mains power quality should be that of a typical
commercial or hospital environment
Mains power quality should be that of a typical
commercial or hospital environment
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
model JC-114N (MH-DR103) image intensifier requires
continued operation during power mains interruption,
it is recommended that the Model JC-114N (MH-DR103)
image intensifier be powered from an uninterruptible
power supply or a battery