d) Lift upward to rotate the upper access door fully into the sealer,
exposing the rollers. When the upper access door assembly clicks
into place, the sealer is ready for cleaning.
3. Clean the rollers:
CAUTION: Do not touch the upper roller if it is hot.
a) Soak up excess water from the drip tray, and then lift up and
remove the drip tray from the bottom of the sealer.
b) To decontaminate the rollers, use a soft cloth moistened with mild
detergent, diluted bleach, or isopropyl alcohol. Use alcohol only if
the roller has cooled completely.
c) To remove debris from the rollers, ensure the roller is cool and dry.
Lightly scrub the roller with a plastic non-scratch scouring pad. Do
not use sharp materials, as this can damage the coating on the
roller. Wipe remaining debris from roller using a soft cloth or paper
towel moistened with mild detergent, diluted bleach, or isopropyl
alcohol.
•
Do not clean or contact the white insulating layer.
•
Roller and/or white insulating layer may develop discoloration
over time, which will not affect sealer performance.
d) Use paper towels or a soft, dry cloth to dry the rollers, interior
surfaces, and drip tray.
e)
To remove debris from sensors A and B, use a cotton swab dipped
in isopropyl alcohol. Sensor B will be silver and shiny when clean.
2d.
3a.
3b.–3d.
A
3e.
Quanti-Tray Sealer PLUS User's Manual
B