Supplied Accessories; Installation - Tecno-gaz Hydra Electronic Plus Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

4. SUPPLIED ACCESSORIES

The autoclave is sold with the accessories included in the article and contained in the same package.
Accessories provided:
N° PIECE
1
Tray extraction and door adjusting key
4
Tray
1
Instruments holder
1
Funnel
1
Tray holder
THE EQUIPMENT LISTED ABOVE DO NOT REQUIRE ADVICES
ATTENTION
LE COUVERCLE DU RESERVOIR D'EAU SE TROUVE A L'INTERIEUR DU KIT
ACCESSORIES

5. INSTALLATION

The installation is a fundamental operation for the following use and good operation of the device.
Here is a list of the points to be followed in order to achieve a good installation.
1. The device must be installed inside a laboratory accessible only to the authorised personnel.
2. The room in which Autoclave is installed must be lighted according to standard UNI 10380 and enough
ventilated.
3. Position the device on a plane surface with minimum capacity of 60 kg.
4. The autoclave feet are adjusted by the producer, in order to obtain a correct level on plane. Their
modifications could compromise the autoclave good operation(see drwg. C).
5. Leave at least 10 cm between the device rear part and the wall.
6. Position the autoclave at such a height as to make it possible for the operator to check the whole
sterilisation chamber and to carry out the normal cleaning operations.
7. Do not install the autoclave near washing basins, taps, because the device furniture is not waterproof.
8. Do not place trays, papers, fluid containers, etc. on the device, because the device grids are used to
ventilate and must not be obstructed.
9. Do not lean on the door when it is opened.
10. Do not install the device near heat sources (for instance: Autoclave, Bunsen burner, kiln); keep a
minimum distance of 40 cm.
11. In case the filling tank is directly discharged in waste waters, position the device at a height which is
higher than the discharge.
12. The device cannot be built in because the water filling in the tank is carried out from above.
13. Install the device in such as way as not to excessively bend the supplying cable.
14. Make sure that the electrical system to which the device is connected is produced according to the laws
in force and has the suitable dimensions for the device features; see PAR.2 of this manual
15. Connect the device pin to a socket with a rated voltage of 230V±10% MONOPHASE + EARTH,
DESCRIPTION
14

Publicidad

Tabla de contenido
loading

Tabla de contenido