11. MAINTENANCE
CAUTION: DISCONNECT ALWAYS THE MAIN SUPPLY BEFORE EACH TECHNICAL INTERVENTION
11.1. DAILY ROUTINE MAINTENANCE
CLEANING THE DOOR SEAL
- Clean the seal on the door and on the outer edge of the sterilisation chamber sealing the gasket
using a wet cloth or the soft part of the supplied sponge.
This cleaning operation must be carried out to remove possible dirt which can cause a pressure loss in the
sterilisation chamber and a possible seal cut.
- Check the water level in the tank( see PAR.8)
11.2. WEEKLY ROUTINE MAINTENANCE
CLEANING THE STERILIZATION CHAMBER
- To remove the deposits on the chamber floor, clean with the abrasive part of the supplied sponge after
removing the tray-holder.
- To moisten the sponge, only use distilled or demineralized water.
- Empty the water tank ( see PAR.9).
11.3. QUARTERLY MAINTENANCE
- Lubricate the hinges and the closing pin with silicon oil.
- Adjusting the door:
If the autoclave handle is too loose while closing, adjust as shown in the following pictures. Rotate in
counterclockwise direction (fig. 11) to increase the seal pressure on the chamber. In case it has loosened too
much and you are not able to close the handle, rotate in direction clockwise (fig. 12).
FIG. 11
FIG. 12
19