Alarms and Alerts
The Toto Touch control unit has three integrated alarms and one information alert to ensure patient
safety and efficacy.
Alarms and Alerts
Air Leak Alarm
The pressure in the system
has fallen below the minimum
operational requirements, i.e.
an air leak.
Blockage Alarm
Airflow between the control unit
and the platform is prevented, i.e. a
blocked air hose.
Service Interval
Service required; default at one-year
use.
Power Loss Alarm
Power to the control unit has been
interrupted.
Initialisation Failure
Control unit fails to start
The sound pressure level of the alarm audio or the reminder audio signal ranges from 55dB-85dB.
The alarm frequency is 150HZ to 1000HZ.
While troubleshooting, the alarm can be muted for up to 10 minutes. If the fault is not corrected
during this time, the alarm reactivates.
If a new alarm condition occurs within 10 minutes, mute will cease, and the new condition will
generate an alarm.
Please Note:
•
The corresponding alarm indication LED remains lit when the alarm is muted.
•
If the alarm condition is corrected while muted, the LED and mute will automatically terminate.
•
The alarm audio signal comes from a speaker, and the remainder audio signal comes from
a buzzer.
•
Alarm activation may take up to two minutes from when the alarm state occurs.
Alarm Condition
Low priority
Low priority
n/a
Low priority
n/a
INFLATE TIME
L & R
R
touch
L
Indicated by an audible alarm and:
Air Leak LED permanently lit yellow
(when activated) (6).
Blockage LED permanently lit yellow
(when activated) (7).
Service LED permanently lit amber
(8) (NO AUDIBLE alarm).
Audible alarm, no LED lit.
All 'Inflate time LEDs' and 'Rest time
LEDs' flashing (when activated).
240 min
180 min
30 min
120 min
60 min
60 min
120 min
30 min
180 min
6
240 min
7
REST TIME
17
5
8