Stimulation Mode - CefarCompex Rehab 400 Manual Del Usuario

Ocultar thumbs Ver también para Rehab 400:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

NOTE: The timer for ch 1+2 will control
ENG
the total session time. This means that
the total time for ch 3+4 could not ex-
ceed the session time for ch 1+2.
NOTE: The program for ch 3+4 is always
in one phase
NOTE: If a preset program uses more
than two channels, it's not avalable for
the 2+2 mode.
The horizontal bars on top of the display
shows the total duration and different
phases of the programs. Left bar for P1
and right bar for P2
P1 shows the program chosen for
channel 1 and 2
P2 shows the program chosen for
channel 3 and 4
F) Stimulation mode
12a
12 b c d e: The different energies
reached during the contraction phase
are shown by a series of black bar
graphs. Active rest phase energies are
shown by hatched bar graphs.
NOTE: Active rest phase stimulation
energies are automatically set at 50%
of contraction intensities but can be
modified during the rest phase. Once
modified, they will be totally independ-
ent of the contraction intensities.
14
12b
12c
12d
12e
Pause mode
12a: Press the On/Off button to inter-
rupt the programme momentarily. To
restart it, simply press the channel 4 +/–
button (12e). The session will resume at
80% of the energy levels that were being
used prior to the interruption.
NOTE: If your Rehab 400 emits a beep-
ing sound and the + symbols under
the active channels begin to flash, the
stimulator is suggesting you to increase
the level of the stimulation energies. If
you are working at the patient's maxi-
mum tolerance level, simply ignore this
message.