TROUBLES SHOOTING
Symptom
Nothing is displayed when
you push the Start button.
"Err P" is shown on LCD
panel
"Err 1" is shown on LCD
panel
"Err 2" is shown on LCD
panel
"Err 3" is shown on LCD
panel
If correct measurement is impossible even after checking the above-mentioned points, consult at the store
•
where you have purchased the unit or the nearest Hyundai dealer without touching the internal mechanism.
• In some very rare cases, there may be error due to the physical condition of the person. In such cases, please
consult your doctor.
All manuals and user guides at all-guides.com
Problem
Are the batteries exhausted?
Are the polarities of batteries in
the wrong position?
Fail to inflate
Deflate too fast
1.Movements occur on the arm
2.Other interference
signal around BPM
Incorrect measured result
Solution
Replace the batteries with new ones.
Insert batteries in the correct direction
1. Please check if the air adaptor on Main Unit and Cuff
are loosen, and connect them.
2. Please check air tube if it is broken and change it.
3. Please check arm cuff If it is broken and change it.
Please contact authorized Hyundai dealer.
1. Keep still after you push the Start button.
2. Please remove interfering sources such as magnetic
material or cell phone.
1. You might wear thick jacket and your BPM fails to
detect pressure values. Please take off your jacket and
measure again.
2. You might use it incorrectly, such as measure
pressure value on child.