Troubleshooting - Bühler technologies BCM-WS Guía Rápida

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 20
BCM

7.1 Troubleshooting

Problem/
Possible cause
malfunction
No display
• No supply
voltage
Error messages on the display:
Alternating between Err and Exxx: e.g.
Error 001
• Ambient tem-
perature too
low
• Ambient tem-
Error 002
perature too
high
Error 004
• Pt100 failure
(short-circuit)
• Pt100 failure
Error 008
(cable break)
Error 032
• Sensor failure
(supply line
open)
• Transmitter
Error 064
(moisture)
Error 128
• Transmitter
(moisture)
• Transmitter
Error 256
(temperature)
Error 512
• Transmitter
(temperature)
Error 1024 • Internal error
Possible errors
Problem /
Possible cause
Malfunction
Switching out-
• Switching out-
put not trig-
put configured
gering when
incorrectly
exceeding lim-
its
• Switching out-
put defect
Switching out-
• Switching out-
put constantly
put configured
switching
incorrectly
• Switching out-
put defect
12
Bühler Technologies GmbH
All manuals and user guides at all-guides.com
Action
The analogue
doesn't re-
• Check cable and
ceive the full/
replace, if ne-
correct output
cessary
current
• Maintain limits
Analogue out-
put doesn't
change the
output signal
• Maintain limits
when the in-
put signal
• Send device in
changes
for repair
• Send device in
for repair
• Send device in
for repair
• Input signal too
low
• Input signal too
high
• Input signal too
low
• Input signal too
high
• Please contact
customer ser-
vice
Action
• In submenu Coux:
"Test Switching
Output" to ensure
normal mode
• In submenu Coux:
"Test Switching
Output" to test the
desired switching
output
• In submenu Coux:
"Test Switching
Output" to ensure
normal mode
• In submenu Coux:
"Test Switching
Output" to test the
desired switching
output
Problem /
Possible cause
Malfunction
• Wrong signal
type set
• Load too high
(current output)
• Analogue output
configured in-
correctly
Action
• In submenu Aoux:
Check and if neces-
sary set the correct
signal type (cur-
rent/voltage out-
put)
• Reduce load to per-
missible value
• In submenu CAnx:
„Test Analogue Out-
put" to ensure nor-
mal mode
BX150101 ◦ 09/2020