Instructions for use
Outlet valve
1.
Slightly press on the valve body with your fingers from the side and remove the cover.
NOTICE
Outlet valves are located on both sides of the mask.
2.
Press a new inlet valve membrane into the inner mask opening.
3.
Hold the membrane stem and turn it to verify its stability under load.
4.
Close the valve body.
Inner mask
1.
Pull the inner mask to remove it.
Insert a new inner mask into the face-piece section according to the drawing, seal the first
2.
one A(a), B(b) and finally C(c), in this order.

Front cover

Use a hexagonal wrench to remove the two screws on the front side and then remove the
1.
front cover.
2.
Use the screws to fasten a new front cover.
Headband
(see Figures Annex 8.1 - 8.4)
1.
Pull the two pawls of the headband from the upper section of the buckle.
2.
To release the buckle turn it on its pin by 90° clockwise and counter-clockwise.
3.
Slide off the buckle to remove the head piece.
4.
Replace the pin.
5.
Install a new headband.
(see Figures Annex 6.1 - 6.4)
(see Figures Annex 7.1 - 7.2)