Setting And Calibration; General Calibration (Cal1) - Baxtran MPI-6 Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ES
FR ENG
6. When the total is shown, pressing
Pressing
numerical keys will exit this mode without clearing the recorded data.
1. Press and hold any key while powering on the scale ON. CAL is shown on the LCD display.
2. Press
to cycle through the calibration menus: CAL 1 and CAL 2.
Select CAL 1, press
Select CAL 2, press
first. Put JP3 back when the calibration is complete).
GENERAL SETTING (CAL 1)
Press
to toggle amount the options, and press
Aut. oFF
1.
- AUTO SHUTDOWN SETTING
DISPLAY
Aut. oFF
0
Aut. oFF
1
Aut. oFF
2
Aut. oFF
3
Aut. oFF
4
bL
2.
- BACKLIGHT SETTING
DISPLAY
b. LigHt
0
b. LigHt
1
b. LigHt
2
b. LigHt
3
User's guide
All manuals and user guides at all-guides.com
will clear all the data on memory.
to enter General Setting menu.
to enter the Calibration Procedure (need to remove the jumper, JP3, from the PCB
EXPLANATION
Disable shutdown function.
If no operation, shutdown the scale in 5 minutes.
If no operation, shutdown the scale in 10 minutes.
If no operation, shutdown the scale in 20 minutes.
If no operation, shutdown the scale in 30 minutes.
EXPLANATION
Shut down backlight
Turn on backlight
Automatic backlight
Manual.

3. SETTING AND CALIBRATION

to confirm your setting.
MPI
15