ENGLISH
DIAGNOSTICS AND TROUBLESHOOTING
To solve the problem, reset the
If the button is pressed
lamp by turning it off and on
The CONTROL
too hard, the lamp may
again. Should the problem persist,
button does not
turn off or not function
recharge the battery.
function correctly
correctly
Press buttons gently
The battery may be
Recharge the battery or remove
flat, or there may be
The display does
it, clean contacts and replace in
problems with the
not turn on
its specific housing
battery connection
The lamp turns off
The lamp is not charged, or the start button has been
during use
accidentally pressed down for more than 2 seconds
Radiation intensity
Cap damaged
Clean or replace the cap
is reduced
If connections have
not been carried out in
The battery does
accordance with the
Charge the battery correctly
not charge
manual, the battery
will not charge