Prepare MLog for a transport
5
Checking and operation of MLog during ongoing operation
(12) Configure the settings for air humidity if it is im-
portant to monitor the humidity for your transport
goods .
(13) Configure the settings for the ambient temperature
if it is important to monitor the temperature for
your transport goods .
(14) Optional:
If your MLog IM100 is equipped with an optio-
nal GSM module, enter the recipient telephone
numbers for sending SMS messages if you would
like to be notified immediately by text about
specific events.
NOTE
Please note that SMS messages can only be sent
with an activated SIM card and this is only possible
from areas with GSM provision . Also take into
account your SIM card's ability to send international
SMS messages .
(15) Optional:
You can now determine how frequently and/or for
what events the MLog should send SMS messa-
ges to the relevant recipient .
Activate SMS messaging per time interval and
configure the time interval and the associated
recipient numbers . The MLog will send an SMS
message during this time period and therefore
enable remote traceability of the transport .
NOTE
Frequent SMS shortens the battery life!
The MLog can also send an SMS message to
specific recipients in the event that limit values are
exceeded. Highlight the appropriate fields in the
software to do this .
(16) After completing all the settings, create a configu-
ration template from these data in order to transfer
it later to a connected and compatible MLog .
This will enable you to use your personal default
settings quickly for several devices .
(17) Now disconnect the link between MLog and
Analyser software using the program function "Cut
Connection to MLog" . Remove the USB cable .
Briefly press the "ON" button on the MLog and pay
attention to the "ACTIVE" LED . It must now light
up for 3 seconds thus showing that the MLog is
ready to record . The LED displays are described
in Section 3 .
26
NOTE
The recording must not start until after the device
has been fitted because the MLog determines the
fitting position as the normal position on starting.
This position is required so that acceleration values
can be determined correctly .
(18) Mounting the MLog on the transport good to be
monitored .
VIDEO
See also video instruction
„MLog-Assembly-Disassembly"
in the MLog Analyser software under
Help menu > Welcome > Introduction Videos .
It is recommended to place the MLog at solid and
non-flexible structures of the shipped equipment
which will not oscillate or vibrate caused by
transport movements . Placement at other
positions may lead to faulty monitoring data and
wrong interpretation in the transport log book and
report . The device should be mounted as close as
possible to the center of gravity; ideally two
devices can be used diagonally opposite at each
end of the equipment . When choosing the
mounting position ensure also that wireless GPS
and GSM communication signals are not blocked
or attenuated by metal structures like cover plates,
for example . Ideally there is a free line of sight to
the sky in a 180 degree sector at least .
(19) The MLog starts the transport recording at the
programmed time . Alternatively, the recording can
be started manually by pressing the "ON" button
(> 4 seconds) .
5
Checking and operation of MLog during
ongoing operation
(1)
Open the MLog's protective cover and
press the "ON" key on the device for about 2
seconds and then again for for about 1 second .
(2)
The LEDs will now indicate the current status of
the device in accordance with Sections 3 .3/3 .5 .
(3)
Replace the batteries if the battery LED flashes
(see Section 3 .7 Battery replacement) .
(4)
Check the device's logbook on the display (only
IM100) or by using the MLog Analyser software if
the "Event/Alarm" LED flashes.
BA3431002/04 ES-EN