System Maintenance

The Microfil funnels and membranes are not reusable and should be discarded
after single use.

Adapting manifold setup for right- or left-handed users

If you need to reconfigure the manifold setup in order to use it more comfortably,
follow the procedure below.
1. Turn manifold tubing adaptor to the side closest to the vacuum source.
2. Change the orientation of the lever by disconnecting the vacuum tubing
attached to the bottom of the stem, unscrewing the wingnut, and rotating the
stem 180°. Then retighten the wing nut and reconnect the tubing.
3. Remove the valve clip with pliers to reposition the vacuum knob on the
opposite side of the stem, with the Millipore logo positioned horizontally. Then
replace clip.