RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND ICARE TA01i
Icare TA01i is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the IcareTA01i can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and Icare TA01i as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
(W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
10
All manuals and user guides at all-guides.com
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
800 MHz to 2,5 GHz
d = 2.3 √P
0.23
0.73
2.3
7.2