Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 47
Protection of the device
Ensure that the substances used
WARNING
with the device are compatible
with the materials of the device
parts that come into contact with
the product; see the chapter "Product
contact parts".
The specified settings on the rating
CAUTION
plate must coincide with the actual
power supply.
The device is only disconnected from the power supply when
the power or device plug is removed.
The socket for the power supply cable must be easy to access.
Covers or parts that can be removed from the device without
tools must later be refitted to ensure safe operation, unless
anything else is connected at this point. This will prevent the
infiltration of foreign objects, liquids and other contaminants.
Unwanted liquids that have entered can be removed by
extracting air during no-load operation.
Avoid knocks or impact to the device.

Unpacking

Unpack the device carefully.
Any damage should be notified immediately to the shipping
agent (post office, railway network or transport company).
Scope of delivery
Diaphragm vacuum pump IKA Vacstar digital
Power supply cable
Laptop Euro
Power supply cable
Laptop GB
Power supply cable
Laptop CH
USB A - USB B cable
Vacuum hose 1m
The device must only be opened by trained, skilled personnel
(who have been authorised to do so).
The device must not be modified without authorisation from
IKA.
To ensure sufficient cooling for the vacuum pump IKA
Vacstar digital, the ventilation slits on the housing must not
be covered.
Only use original replacement parts for maintenance to
ensure that the device works reliably.
Watch out for water condensation inside and outside the
device. Warm up the device first, if it has been brought in
from a cold environment.
Never fasten the vaccum pump over a heating bath.
Ensure that no solids and/or liquids can enter the diaphragm
of the vacuum pump IKA Vacstar digital through the suction
line of the pump. This damages the diaphragm and other
internal parts of the pump. This reduces delivery capacity and
the final pressure can no longer be achieved. Deposits may
be produced in the interior, reducing service life and causing
leaks.
Unpacking
Silencer
Connecting cable MVP
10.100
Operating instructions
Warranty card
21

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

20000016633

Tabla de contenido