Error Messages - Fora Test N'GO Advance Pro Manual Del Propietário

Sistema de monitoreo multifuncional (glucosa sanguínea y b-cetona)
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Error Messages

MESSAGE
E-b
E-2
E-U
E-E
E-0
E-A
E-F
E-C
E-t
32
WHAT IT MEANS
Appears when the batteries
cannot provide enough
power for a test.
Expired code strip.
Appears when a used test
strip is inserted.
Problem in operation.
Appears when test strip is
removed while counting
down.
Appears when the wrong
code strip is inserted or
other coding errors.
Appears when ambient
temperature is below or
above system operation
range.
WHAT TO DO
Replace the batteries
immediately.
Repeat the test with a new
lot of test strip.
Repeat with a new test strip.
Repeat the test with a new
test strip. If the meter still
does not work, please
contact the customer service
for assistance.
Repeat the test with a new
test strip.
Make sure to insert the right
code strip for the meter.
Check if the code numbers
on the code strip and the
strip vial label are the same.
System operation range is
46.4°F to 113°F (8°C to
45°C). Repeat the test after
the meter and test strip are
in the above temperature
range.