A light bar will indicate the menu choice. As the menu is scrolled through, the light
bar will highlight different menu choices. Pressing the
menu choice that is indicated by the light bar.
In the second format the menu choice window takes advantage of the graphical
capabilities of the display. Large format graphic information, such as test results
or error messages or the LaMotte logo is displayed. The top two lines of the display
are used to display information in a large, easy to read format. The menus work in
the same way as previously described but two lines of the menu are visible at the
bottom of the display.
Message or Result Window
Main Window Selection
As described previously, the
menu and a return to the previous menu. This allows a rapid exit from an inner menu
to the main menu by repeatedly pushing the
time will turn the colorimeter off.
The display may show the following messages:
Battery Status
More choices are available and can be viewed by scrolling up and/or
down through the display.
Header
Identifies the current menu and information on units and reagent
systems if applicable.
Footer
In the data logging mode the number of the data point is displayed
and the total number of data points in the memory will be shown. The
footer also shows current time and battery status
TUBES AND CHAMBERS
The colorimeter uses one type of tube (Code 0290) for all test factors.
The handling of the tubes is of utmost importance. Tubes must be clean and free
from lint, fingerprints, dried spills and significant scratches, especially the central
zone between the bottom and the sample line.
Scratches, fingerprints and water droplets on the tube can cause stray light
interference leading to inaccurate results. Tubes that have been scratched in the
light zone through excessive use should be discarded and replaced with new ones.
Tubes should always be washed on the inside and outside with mild detergent prior
to use to remove dirt or fingerprints. The tubes should be allowed to air-dry in an
inverted position to prevent dust from entering the tubes. Dry tubes should be
Header