Accuracy Of Meter Measurements - Prodigy AutoCode Manual Del Propietário

Ocultar thumbs Ver también para AutoCode:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

PRODIGY AutoCode® Blood Glucose Meter

Accuracy of Meter Measurements

FAQ: Can two (2) blood glucose meters produce
different readings that are from the same blood
sample? Does this make one reading more
accurate than the other?
Answer: Yes, different meters can produce
different readings using the same blood
sample.
No, this does not necessarily mean that
one result is more accurate than the
other.
The reasons for this are as follows:
1. The FDA recognized standard for glucose
meter accuracy is ISO 15197: In Vitro
Diagnostic Test Systems Requirements
for Blood Glucose Monitoring Systems for
Self Testing in Managing Diabetes Mellitus.
This standard sets accuracy requirements
that manufacturers must meet before being
approved by the FDA to market their
glucose meters.
Owner's Manual
PRODIGY AutoCode® Blood Glucose Meter
2. The accuracy requirements for the blood
glucose monitoring system—meters and test
strips—as stated in ISO 15197 are:
Ninety-five percent (95%) of the individual
glucose results shall fall within ±15 mg/
dL (0.83 mmol/L) of the results of the
manufacturer's measurement procedure
at glucose concentrations < 75 mg/dL
(<4.2 mmol/L) and within ± 20% at glucose
concentrations ≥ 75 mg/dL ( ≥ 4.2 mmol/L).
How does this answer the FAQ
The accuracy of the system is determined by a
clinical study using 100 blood samples at interval
glucose concentrations ranging from < 50 mg/
dL to over 400 mg/dL measured using both
the manufacturer's meter and a clinical blood
glucose analyzer for comparison (reference
measurement).
Example
If the manufacturer's reference measurement is 70
mg/dL this means that the meter measurements
are considered accurate if they fall within ± 15 mg/
dL of 70 mg/dL (from 55 mg/dL to 85 mg/dL).
70
Owner's Manual
71

Publicidad

Capítulos

Tabla de contenido
loading

Productos relacionados para Prodigy AutoCode

Tabla de contenido