Function Check; Troubleshooting - Seca 644 Manual De Instrucciones Y Garantía

Ocultar thumbs Ver también para 644:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 231

8. FUNCTION CHECK

9. TROUBLESHOOTING

Problem
No weight displayed
with a load on the scale.
0.00 does not appear
before weighing.
One segment is illumi-
nated either continu-
ously or not at all.
The display
appears.
The display "bAtt"
appears.
Perform a function check before each use.
A complete function check includes:
• a visual inspection for mechanical damage
• a visual inspection and function check of the display
• function check of all the controls shown in the
section entitled "Overview" on page 71
• a function check of the optional accessories
If you find faults or deviations during the function
check, as a first step try to remedy the fault with the aid
of the section entitled "Troubleshooting" from
page 109.
CAUTION!
Bodily injury
If you find faults or deviations during the func-
tion check which you are unable to remedy with
the aid of the section entitled "Troubleshooting"
from page 109, you may not use the device.
Have servicing and repairs carried out
exclusively by seca service or by an
authorized service partner.
Please see the section entitled "Servicing" on
page 111.
Cause / Solution
The scale has no power supply.
- Check whether the scale is switched on
A load was already on the scale before it was
switched on.
- Remove the load from the scale
- Switch the scale off and back on again
The corresponding point has a fault.
- Inform the service department
Battery block power is running down.
- Recharge battery block as soon as possible
Battery block discharged.
- Charge battery block
English
Function check • 109

Publicidad

Tabla de contenido
loading

Tabla de contenido