Bühler technologies EGK 4S Guía Rápida página 11

Compresor-refrigerador de gases de muestreo
Ocultar thumbs Ver también para EGK 4S:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 17
EGK 4S
No cooling
• Compressor
doesn't start up
Fuse is trigger-
• Increased com-
ing
pressor current
consumption due
to faulty com-
pressor start-up
Error Messages in the Display
The display alternates between the temperature and error
message,
Problem / Mal-
Possible cause
function
Error 01
• Interruption
Error 02
• Short circuit
BX450004 ◦ 07/2019
• Compressor PTC
not cooled down
sufficiently. Wait
5 minutes and try
again.
• Compressor PTC
not cooled down
sufficiently. Wait
5 minutes and try
again.
Action
• Temperature
sensor failure:
Send in cooler
• Temperature
sensor failure:
Send in cooler