Differential Pressure Display - WIKA 712.15.160 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

4. Commissioning, operation
Panel mounting (option)

4.2 Differential pressure display

The measuring span of the differential pressure gauge can, depending on the measuring
cell, be adjusted within the measuring range limits given in the table. Ideally, this adjustment
should be made on a test bench, though it can also be carried out at the measuring point
using a hand test pump.
Measuring range limits
Measuring cell Adjustable scale ranges
from
140 mbar
0 ... 40 mbar
280 mbar
0 ... 80 mbar
560 mbar
0 ... 160 mbar
1,130 mbar
0 ... 320 mbar
2,300 mbar
0 ... 650 mbar
4,000 mbar
0 ... 1,150 mbar
Differential pressure gauge models 712.15.160, 732.15.160, optional for hazardous areas
to
0 ... 140 mbar
0 ... 280 mbar
0 ... 560 mbar
0 ... 1,130 mbar
0 ... 2,300 mbar
0 ... 4,000 mbar
Panel cutout
Panel
EN
13

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

732.15.160

Tabla de contenido