6.2.
Switch-over from warning to shutdown function
Factory setting:
>WARNING<
Safety installations, warning functions
If a shutdown of functional elements (e.g. heater, circulating pump) is required when
the limit values are exceeded or undercut the circulator can be changed over from
warning function >WARNING< to shutdown function >ALARM<.
1. Press the key
2. Select the menu >SECVAL -AL-TYPE< by pressing the
3. Press the
key and the set parameter will flash
(Example: WARNING)
4. Change the parameter by pressing the
key.
or
press the
key if the parameter is to retained.
Setting >WARNING<
A mere warning function with optical and audible warning signal (interval tone) A
message appears on the VFD COMFORT-DISPLAY:
or
OVERTMP
• Setting >ALARM<
Temperature limit with shutdown of heater and circulating pump.
An audible alarm sounds (continuous tone) and a message appears on the VFD
COMFORT-DISPLAY:
or
OVERTMP
.
SUBTEMP
SUBTEMP
75
key.
.
key and confirm by pressing the