Xylem SI Analytics CT 72/2 Manual De Instrucciones página 99

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

in commands: Asking for parameters or temperature values to be displayed.
Command
version
status
in_pv_00
in_pv_01
in_pv_03
in_pv_04
in_sp_01
in_sp_02
in_sp_03
in_par_01
in_par_02
in_par_03
in_par_05
in_par_06
in_par_07
in_par_08
in_mode_01
in_mode_05
in_mode_08
Parameter
Response of circulator
none
Number of software version (V X.xx)
none
Status message, error message (see page 100)
none
Actual bath temperature.
none
Heating power being used (%).
none
Temperature value registered by the safety sensor.
none
Setpoint temperature of the excess temperature protection
none
Working temperature „SETPNT 1"
none
High temperature warning limit „OverTemp"
none
Low temperature warning limit „SubTemp"
none
Te - Time constant of the external bath.
none
Si - Internal slope
none
Ti - Time constant of the internal bath.
none
Factor pk/ph0:
Ratio of max. cooling capacity versus max. heating capacity
none
Xp control parameter of the internal controller.
none
Tn control parameter of the internal controller.
none
Tv control parameter of the internal controller.
none
Selected setpoint:
0 = SETPNT 1
1 = SETPNT 2
2 = SETPNT 3
none
Circulator in Stop/Start condition:
0 = Stop
1 = Start
none
Adjusted control dynamics
0 = aperiodic
1 = standard
Remote control
99

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Si analytics ct 72/4Si analytics ct 72/2 tt

Tabla de contenido