Menu Limits; Limits For Internal Control - Xylem SI Analytics CT 72/2 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

7.7.

MENU LIMITS

Menu- Level 1
Level 2

7.7.1. Limits for internal control

Press the
key if parameter is to be retained. Correction function for
parameters or values (prior to ENTER).
Parameter level
0 ... 100 %
0 ... 100 %
SETPOINT MAX / MIN – Maximum and minimum setpoint
Restriction of the adjustable temperature range
The limitation of the operating temperature range effects the temperature setting
in the menu with the key
Only setting of working temperatures which lie within the determined limits is
possible
Existing settings for SETPNT 1, -2, -3, as well as those for >OVERTMP< and >
SUBTMP < (refer to page 75), are automatically deferred into the limit range.
Setting range: -94,90 °C ... +200,0 °C
SET MAX > SET MIN
Interchange of values is not possible.
Set maximum heating / cooling
The heating and cooling capacity of the unit are adjustable. 100 % corresponds to
the technical specification of the equipment.
Setting range:
HEAT MAX –
COOLING MAX –
Menu functions
Integer digits flash, set by pressing
+
Decimal digits flash, set by pressing
+
Integer digits flash, set by pressing
+
Decimal digits flash, set by pressing
+
The value flashes, set by pressing
+
The value flashes, set by pressing mit
+
.
0 to 100 % in 1 % steps
0 to 100 % in 1 % steps
93

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Si analytics ct 72/4Si analytics ct 72/2 tt

Tabla de contenido