and then screw the gas purge valve (with a torque
of 2.5 Nm) as shown in the following figure.
VENT VALVE INSTALLATION
The vent valve mod. 969-9833 allows to avoid un-
desired venting of the pump during a temporary
power failure (5 sec maximum), and enables an
automatic vent operation.
Unscrew the threaded plug (see figure below).
All manuals and user guides at all-guides.com
31
TECHNICAL INFORMATION
Screw the vent valve into the pump without using
the supplied adapter.
Tighten the valve using a 16 mm. hexagonal span-
ner with a torque of 2.5 Nm.
! ! ! !
CAUTION
Do not overtighten the valve as this may damage
the thread on the pump.
87-900-891-01(C)