Each cleaning and disinfection cycle includes a pre-cleaning step with
one wipe and a disinfection step with a second wipe. Improper system
cleaning and disinfection may result in meter malfunction. If you have
a question, please contact customer service at 1-713-842-7252 for
assistance.
This device has been validated to withstand 260 cleaning and
disinfection cycles using the recommended disinfecting wipe/towelette.
The tested number of cycles is estimated by 1 cleaning and disinfection
cycle per meter per week for 5 years, the expect life of device.
Stop using the meter if you see any signs of deterioration such as the LCD
display is cracked or cloudy, the buttons no longer function, or the outer
casing is cracked. Please contact customer service at 1-713-842-7252
for a replacement meter if any of the signs of deterioration are noticed.
The meter should be replaced after the validated number of disinfection
cycles or the warranty period, which ever comes first.
NOTE :
• Do NOT clean and disinfect the meter while performing tests.
• If the meter is being operated by a second person, the meter and lancing
device should be decontaminated prior to use by the second person.
• Do NOT allow cleaning and disinfecting solution to get in the test slot, the
battery compartment or the eject button.
• If you do get moisture in the test strip slot, wipe it away with a corner of tissue.
• Always dry the meter thoroughly before using it.
• Do not spray the meter directly with cleaning solutions especially those
containing water (i.e. soapy water), as this could cause the solution to enter
the case inside and damage the electronic components or circuitry.