PIPMIX MODE The Programs · Settings and Use
Program for mixing of liquids. The sample is repeatedly aspirated and discharged.
Volume and speed adjustments are described on pages 8 and 9.
What to do
1.
Bring up the menu
2.
Select PIPmix
mode
3.
Confirm PIPmix
mode
4.
Prepare for
pipetting
5.
Aspirate liquid
14
· ErgoOne E -8/-12 · User Manual
How to do it
Press the MENU key three
times to bring up the mode
selection menu. 'Mode'
blinks.
Scroll through the modes us-
ing either the UP or DOWN
key until 'PIPmix' appears.
'Mode' continues to blink.
Press the ENTER key. The
Display now shows 'blo' for
blow-out.
Press the pipetting key once
to move the pistons into the
start position.
The arrow in the display
points upwards (aspiration).
Press the pipetting key once
to aspirate the liquid.
Keys to press
Display readout
3x
1x
1x
1x