6.
Place the new membrane filter with
the thicker end facing downwards into the
adapter (with the smaller conical side up).
7.
Screw in the adapter support hand tight.
8.
The entire instrument incl. membrane
filter is autoclavable at 121 °C (2 bar)
according to DIN EN 285.
For autoclaving remove suction bellow.
Note:
Not every membrane filter is suitable. Only use genuine manufacturer's
recommended accessories. Filters of a pore size of 3 µm are a standard
equipment of the instrument. The membrane filter is autoclavable up
to 5 times at 121 °C. It is the user's responsibility to ensure effective
autoclaving.
Cleaning · Autoclaving