Automated Cleaning
3.Cleaning.was.validated.using.
Enzol®.at.1.oz/gal.
4.Cleaning.was.validated.using.
Renu-Klenz™.at.¼.oz/gal.
EN-6
1
•. Disassemble.the.tray.into.its.individual.
•. Rinse.the.tray.with.RO/DI.water.at.ambient.
•. Place.the.tray.in.the.washer.on.an.incline.to.
2
•. Program.the.washer.using.the.settings.below.
Recirculation
Pre Wash
Enzyme
Wash
Wash 1
Rinse 1
Dry Phase
3
•. Remove.the.tray.from.the.washer.after.the.
•. Dry.the.tray.using.a.clean.soft.cloth..Use.filtered.
•. Visually.inspect.the.tray.for.cleanliness,.paying.
Rinse
components:.the.base,.lid,.and.(if.applicable).
silicon.mat.
temperature.until.there.is.no.visible.detergent.
residue..Use.a.syringe.to.assist.in.rinsing..After.
all.detergent.residues.have.been.removed,.
continue.to.rinse.for.a.minimum.of.30.seconds.
facilitate.drainage.
Automated wash
Time
Temperature
2.min.
Cold.tap.water
2.min.
Hot.tap.water
2.min.
Set.point.60°C
2.min.
7.min.
Drying Temperature
Drying
completion.of.the.dry.phase.
pressurized.air.(40.psi).to.assist.in.drying.
close.attention.to.hard-to-reach.areas.
Water
(140°F)
Heated.60°C
(140°F)
115°C.(239°F)
Detergent
(type and
concentration)
—
Enzymatic.
Detergent3
Non-enzymatic.
Detergent4
—
—