Cable Connections; General Information; Connecting Cables - Minebea Intec PR 6130/08 Manual De Instalación

Ocultar thumbs Ver también para PR 6130/08:
Tabla de contenido

Publicidad

Cable Junction Box PR 6130/08
5

Cable connections

5.1

General information

5.2

Connecting cables

5.3
Cable connections
EN-14
Cable entry must be from below.
Fit the wires and screen of the connection cable with wire end ferrules as per DIN 46228-1:
2
-
0.5 mm
for green, gray, black, and white
2
-
0.75 mm
for blue and red
This is not necessary for the load cell cables.
Connect the wires to the terminals according to the color coding .
Connect the screens on the other side of the connection cable with the equipotential
bonding terminal of the downstream device (see instrument manual) as described in
Chapters
5.3
and 5.4.
To connect the junction box to the device, connection cable PR 6135/.. must be used;
max. length: 300 m.
Note:
All components are only shown schematically.
Color code
bk
=
black
bu
=
blue
gn
=
green
gy
=
gray
rd
=
red
wh
=
white
5 Cable connections
Minebea Intec

Publicidad

Tabla de contenido
loading

Tabla de contenido