Important Notices; Error Codes - LaserLiner LaserRange-Master T4 Pro Manual Del Usuario

Ocultar thumbs Ver también para LaserRange-Master T4 Pro:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 44

Important notices

– The laser points to the location that will be measured.
No objects may get into the laser's line of measurement.
– The device compensates the measurement for different room
temperatures. Therefore allow the device a brief adaptation period
when changing locations with large temperature differences.
– The device is only conditionally useable in outdoor areas and cannot
be used in strong sunlight.
– The measurement results of outdoor measurements may be influenced
or falsified by rain, fog and snow.
– In unfavourable conditions, e.g. with poorly reflecting surfaces, the
maximum deviation may be greater than 3 mm.
– Carpeting, upholstery or curtains will not reflect the laser optimally.
Measure to flat surfaces.
– Measurements made through glass (window panes) can falsify
measurement results.
– An energy-saving function switches the device off automatically.
– Clean with a soft cloth. Water may not be allowed to penetrate
the housing.

Error codes:

Err204:
Calculation error
Err208:
Internal fault
Err220:
Replace the battery
Err252:
Temperature is
too high: > 40°C
Err253:
Temperature is
too low: < 0°C
Information on maintenance and care
Clean all components with a damp cloth and do not use cleaning agents,
scouring agents and solvents. Remove the battery(ies) before storing for
longer periods. Store the device in a clean and dry place.
Calibration
The measuring device should be calibrated and tested on a regular basis
to ensure it is accurate and working properly. We recommend the me-
asuring device is calibrated every 1 - 2 years. If necessary, contact your
distributor or the UMAREX-LASERLINER service department.
18
EN
Err255:
Received signal too weak
or measuring time too long
Err256:
Received signal too strong
Err261:
Outside the measuring
range
Err500:
Hardware fault

Publicidad

Tabla de contenido
loading

Tabla de contenido